Effect of Leak and Breathing Pattern on the Accuracy of Tidal Volume Estimation by Commercial Home Ventilators: A Bench Study

Fecha de publicación:

Autores de I3PT

Participantes ajenos a I3PT

  • Sales, B

Grupos de Investigación

Abstract

BACKGROUND: New home ventilators are able to provide clinicians data of interest through built-in software. Monitoring of tidal volume (V-T) is a key point in the assessment of the efficacy of home mechanical ventilation. OBJECTIVE: To assess the reliability of the V-T provided by 5 ventilators in a bench test. METHODS: Five commercial ventilators from 4 different manufacturers were tested in pressure support mode with the help of a breathing simulator under different conditions of mechanical respiratory pattern, inflation pressure, and intentional leakage. Values provided by the built-in software of each ventilator were compared breath to breath with the V-T monitored through an external pneumotachograph. Ten breaths for each condition were compared for every tested situation. RESULTS: All tested ventilators underestimated V-T (ranges of -21.7 mL to -83.5 mL, which corresponded to -3.6% to -14.7% of the externally measured V-T). A direct relationship between leak and underestimation was found in 4 ventilators, with higher underestimations of the V-T when the leakage increased, ranging between -2.27% and -5.42% for each 10 L/min increase in the leakage. A ventilator that included an algorithm that computes the pressure loss through the tube as a function of the flow exiting the ventilator had the minimal effect of leaks on the estimation of V-T (0.3%). In 3 ventilators the underestimation was also influenced by mechanical pattern (lower underestimation with restrictive, and higher with obstructive). CONCLUSIONS: The inclusion of algorithms that calculate the pressure loss as a function of the flow exiting the ventilator in commercial models may increase the reliability of V-T estimation.

Datos de la publicación

ISSN/ISSNe:
0020-1324, 1943-3654

Respiratory Care  SAGE PUBLICATIONS INC

Tipo:
Article
Páginas:
770-777
PubMed:
23051878

Citas Recibidas en Web of Science: 57

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • noninvasive ventilation; tidal volume; leaks

Campos de Estudio

Cita

Compartir