Bioreactance is not reliable for estimating cardiac output and the effects of passive leg raising in critically ill patients
Fecha de publicación:
Autores de I3PT
Participantes ajenos a I3PT
- Kupersztych-Hagege, E
- Teboul, JL
- Talbot, A
- Sabatier, C
- Richard, C
- Monnet, X
Grupos de Investigación
Abstract
Bioreactance estimates cardiac output in a non-invasive way. We evaluated the ability of a bioreactance device (NICOM) to estimate cardiac index (CI) and to track relative changes induced by volume expansion. In 48 critically ill patients, we measured CI estimated by the NICOM device (CINicom) and by transpulmonary thermodilution (CItd, PiCCO(2) device) before and after a 500 ml saline infusion. Before volume expansion, we performed a passive leg raising (PLR) test and measured the changes it induced in CINicom and in pulse contour analysis-derived CI. Considering the values recorded before PLR and before and after volume expansion (n144), the bias (lower and upper limits of agreement) between CItd and CINicom was 0.9 (2.2 to 4.1) litre min(1) m(2). The percentage error was 82. There was no significant correlation between the changes in CItd and CINicom induced by volume expansion (P0.24). An increase in CI estimated by pulse contour analysis 9 during the PLR test predicted fluid responsiveness with a sensitivity of 84 (95 confidence interval 6097) and a specificity of 97 (95 confidence interval 82100). The area under the receiver operating characteristic curve constructed to test the ability of the PLR-induced changes in CINicom in predicting fluid responsiveness did not differ significantly from 0.5 (P0.77). The NICOM device cannot accurately estimate the cardiac output in critically ill patients. Moreover, it could not predict fluid responsiveness through the PLR test.
Datos de la publicación
- ISSN/ISSNe:
- 0007-0912, 1471-6771
- Tipo:
- Article
- Páginas:
- 961-966
- DOI:
- 10.1093/bja/aet282
- PubMed:
- 23985531
BRITISH JOURNAL OF ANAESTHESIA ELSEVIER SCI LTD
Citas Recibidas en Web of Science: 91
Documentos
- No hay documentos
Filiaciones
Keywords
- equipment; monitors; dobutamine; measurement; cardiac output; measurement techniques; shock
Campos de Estudio
Cita
Kupersztych E,Teboul JL,Artigas A,Talbot A,Sabatier C,Richard C,Monnet X. Bioreactance is not reliable for estimating cardiac output and the effects of passive leg raising in critically ill patients. Br. J. Anaesth. 2013. 111. (6):p. 961-966. DOI: 10.1093/bja/aet282.