Efficiency of a mechanical device in controlling tracheal cuff pressure in intubated critically ill patients: a randomized controlled study

Fecha de publicación:

Autores de I3PT

  • Paula Saludes Orduña

    Autor

  • Jordi Vallès Daunis

    Autor

  • Ignacio Esteban Martin-Loeches Carrondo

    Autor

Participantes ajenos a I3PT

  • Nseir, S
  • Rodriguez, A
  • De Jonckheere, J

Grupos de Investigación

Abstract

Background: Cuff pressure (P-cuff) control is mandatory to avoid leakage of oral secretions passing the tracheal tube and tracheal ischemia. The aim of the present trial was to determine the efficacy of a mechanical device (PressureEasy (R)) in the continuous control of Pcuff in patients intubated with polyvinyl chloride (PVC)-cuffed tracheal tubes, compared with routine care using a manometer. Methods: This is a prospective, randomized, controlled, cross-over study. All patients requiring intubation with a predicted duration of mechanical ventilation >= 48 h were eligible. Eighteen patients randomly received continuous control of Pcuff with PressureEasy (R) device for 24 h, followed by discontinuous control (every 4 h) with a manual manometer for 24 h, or vice versa. Pcuff and airway pressure were continuously recorded. Pcuff target was 25 cmH(2)O during the two periods. Results: The percentage of time spent with P-cuff 20-30 cmH(2)O (median (IQR) 34 % (17-57) versus 50 % (35-64), p = 0.184) and the percentage of time spent with P-cuff <20 cmH(2)O (23 % (5-63) versus 43 % (16-60), p = 0.5) were similar during continuous control of Pcuff and routine care, respectively. However, the percentage of time spent with P-cuff >30 cmH(2)O was significantly higher during continuous control compared with routine care of tracheal cuff (26 % (14-39) versus 7 % (1-18), p = 0.002). No significant difference was found in Pcuff (25 (18-28) versus 21 (18-26), p = 0.17), mean airway pressure (14 (10-17) versus 14 (11-16), p = 0.679), or coefficient of variation of Pcuff (19 % (11-26) versus 20 % (11-25), p = 0.679) during continuous control compared with routine care of tracheal cuff, respectively. Conclusions: PressureEasy (R) did not demonstrate a better control of P-cuff between 20 and 30 cmH(2)O, compared with routine care using a manometer. Moreover, the device use resulted in significantly higher time spent with overinflation of tracheal cuff, which might increase the risk for tracheal ischemic lesions.

Datos de la publicación

ISSN/ISSNe:
2110-5820, 2110-5820

Annals of Intensive Care  ELSEVIER

Tipo:
Article
Páginas:
-
PubMed:
26059205

Citas Recibidas en Web of Science: 12

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • Cuff pressure; Control; Tracheal ischemia; Microaspiration; Mechanical ventilation