Efficiency of a mechanical device in controlling tracheal cuff pressure in intubated critically ill patients: a randomized controlled study
Fecha de publicación:
Autores de I3PT
Participantes ajenos a I3PT
- Nseir, S
- Rodriguez, A
- De Jonckheere, J
Grupos de Investigación
Abstract
Background: Cuff pressure (P-cuff) control is mandatory to avoid leakage of oral secretions passing the tracheal tube and tracheal ischemia. The aim of the present trial was to determine the efficacy of a mechanical device (PressureEasy (R)) in the continuous control of Pcuff in patients intubated with polyvinyl chloride (PVC)-cuffed tracheal tubes, compared with routine care using a manometer. Methods: This is a prospective, randomized, controlled, cross-over study. All patients requiring intubation with a predicted duration of mechanical ventilation >= 48 h were eligible. Eighteen patients randomly received continuous control of Pcuff with PressureEasy (R) device for 24 h, followed by discontinuous control (every 4 h) with a manual manometer for 24 h, or vice versa. Pcuff and airway pressure were continuously recorded. Pcuff target was 25 cmH(2)O during the two periods. Results: The percentage of time spent with P-cuff 20-30 cmH(2)O (median (IQR) 34 % (17-57) versus 50 % (35-64), p = 0.184) and the percentage of time spent with P-cuff <20 cmH(2)O (23 % (5-63) versus 43 % (16-60), p = 0.5) were similar during continuous control of Pcuff and routine care, respectively. However, the percentage of time spent with P-cuff >30 cmH(2)O was significantly higher during continuous control compared with routine care of tracheal cuff (26 % (14-39) versus 7 % (1-18), p = 0.002). No significant difference was found in Pcuff (25 (18-28) versus 21 (18-26), p = 0.17), mean airway pressure (14 (10-17) versus 14 (11-16), p = 0.679), or coefficient of variation of Pcuff (19 % (11-26) versus 20 % (11-25), p = 0.679) during continuous control compared with routine care of tracheal cuff, respectively. Conclusions: PressureEasy (R) did not demonstrate a better control of P-cuff between 20 and 30 cmH(2)O, compared with routine care using a manometer. Moreover, the device use resulted in significantly higher time spent with overinflation of tracheal cuff, which might increase the risk for tracheal ischemic lesions.
Datos de la publicación
- ISSN/ISSNe:
- 2110-5820, 2110-5820
- Tipo:
- Article
- Páginas:
- -
- PubMed:
- 26059205
Annals of Intensive Care ELSEVIER
Citas Recibidas en Web of Science: 12
Documentos
- No hay documentos
Filiaciones
Keywords
- Cuff pressure; Control; Tracheal ischemia; Microaspiration; Mechanical ventilation
Campos de Estudio
Cita
Nseir S,Rodriguez A,Saludes P,De Jonckheere J,Valles J,Artigas A,Martin I. Efficiency of a mechanical device in controlling tracheal cuff pressure in intubated critically ill patients: a randomized controlled study. Ann. Intensive Care. 2015. 5. 12. DOI: 10.1186/s13613-015-0054-z.