Titration of Mechanical Insufflation-Exsufflation Optimal Pressure Combinations in Neuromuscular Diseases by Flow/Pressure Waveform Analysis
Autores de I3PT
Grupos de Investigación
Abstract
Introduction: The aim of this study was to assess several air-pressure settings for MI-E to determine their effect on peak cough flow (PCF), and to compare the best pressures with those are more common used in the literature (+/- 40 cmH(2)O) in patients with neuromuscular disorders (NMD). Methods: Adults with NMD in whom MI-E was indicated were recruited. Assisted PCF was measured by an external pneumotachograph. The protocol included 9 PCF measures per patient: 1 baseline (non-assisted), 4 with increasing inspiratory pressures without negative pressure (10, 20, 30 and 40 cmH(2)O or maximum tolerated), and then 4 adding expiratory pressures (-10, -20, -30 and -40 cmH(2)O or maximum tolerated) with maximum inspiratory pressure previously achieved. Results: Twenty one patients were included, 61% with amyotrophic lateral sclerosis (ALS). Mean PCFs with recommended pressures (+/- 40 cmH(2)O) were lower than the scored in the individualized steps of the titration protocol (197 +/- 7 671/min vs 214 +/- 2 601/min, p < 0.05). Regarding subgroups, mean PCFmax values in ALS patients with bulbar symptoms were significantly higher than those achieved with recommended pressures (163 +/- 6 80 vs 189 +/- 661/min, p < 0.05). Conclusion: The PCFmax obtained with the protocol did not always match the recommended settings. It may be advisable to perform MI-E titration assessed by non-invasive PCF monitoring in patients with NMD, especially in ALS with bulbar involvement to improve the therapy detecting airway collapse induced by high pressures. (C) 2018 SEPAR. Published by Elsevier Espana, S.L.U. All rights reserved.
Datos de la publicación
- ISSN/ISSNe:
- 0300-2896, 1579-2129
- Tipo:
- Article
- Páginas:
- 246-251
- PubMed:
- 30598235
ARCHIVOS DE BRONCONEUMOLOGIA ELSEVIER ESPANA SLU
Citas Recibidas en Web of Science: 20
Documentos
- No hay documentos
Filiaciones
Keywords
- Mechanical assisted cough; Neuromuscular disorders; Monitor; Flow/pressure waveforms
Campos de Estudio
Cita
Lalmolda C,Prados H,Mateu G,Noray M,Pomares X,Lujan M. Titration of Mechanical Insufflation-Exsufflation Optimal Pressure Combinations in Neuromuscular Diseases by Flow/Pressure Waveform Analysis. Arch. Bronconeumol. 2019. 55. (5):p. 246-251. DOI: 10.1016/j.arbres.2018.10.011.
Portal de investigación