A customized percutaneous three-dimensional-printed guide for scaphoid fixation versus a freehand technique: a comparative study
Autores de I3PT
Grupos de Investigación
Abstract
The aim of this study was to compare the accuracy and reliability of percutaneous fixation of minimally displaced scaphoid fractures using a customized three-dimensional (3-D)-printed guide with a conventional freehand method. A prospective cohort of ten patients underwent scaphoid fixation with the aid of a customized 3-D-printed guide. The final screw position, total surgery time (minutes) and fluoroscopy time (seconds) was compared with a retrospective cohort of ten patients who underwent fixation with a conventional technique. There were no differences in final screw position between both methods. The patients in which the 3-D guide was used had a surgery time reduction of 43% and a fluoroscopy time reduction of 52% compared with the control freehand group. The use of a customized 3-D-printed guide permits a fixation that is as accurate as the standard freehand technique, with reduction in surgical time and intraoperative radiation exposure.
Datos de la publicación
- ISSN/ISSNe:
- 1753-1934, 2043-6289
- Tipo:
- Article
- Páginas:
- 1081-1087
- PubMed:
- 34647500
Journal of Hand Surgery-European Volume SAGE PUBLICATIONS LTD
Citas Recibidas en Web of Science: 7
Documentos
- No hay documentos
Filiaciones
Keywords
- Scaphoid fracture; 3-D-printing; scaphoid fixation; external guide
Campos de Estudio
Cita
Marcano FA,Berenguer A,Fillat F,Corderch S,Camara J,Sanchez R. A customized percutaneous three-dimensional-printed guide for scaphoid fixation versus a freehand technique: a comparative study. J. Hand Surg.-Eur. Vol. 2021. 46. (10):p. 1081-1087. DOI: 10.1177/17531934211049132.
Portal de investigación